C.A.P. REAGENT GRADE WATER STANDARDS
|
PROPERTY
|
TYPE I
|
TYPE II
|
TYPE III
|
|
RESISTIVITY
(@ 25 C)
- In-Line
- Effluent
|
10.0 MOhm/cm
N/A
|
N/A
2.0 MOhm/cm
|
N/A
1.0 MOhm/cm
|
|
TOTAL SILICA
|
0.05 mg/l
|
0.10 mg/l
|
1.00 mg/l
|
|
pH (@25 C)
|
N/A
|
N/A
|
5.0 8.0
|
|
MICROBIAL
|
10 CFU/ml
|
1000 CFU/ml
|
N/A
|
NOTES:
- PARTICULATE MATTER: TYPE I water should be contain less than 500
particles per liter that are greater than 0.20 micron in size.
- Other parameters such as particle size, heavy metal content, and/or
trace elements and carbon dioxide should be considered and monitored
when appropriate for special or specific uses.
- Organisc and other contaminants may be reduced with the use of
activated carbon.
- Any method of preparation of the three types of water is acceptable